摘要: |
利用AFS-820双道原子荧光光度计对水中微量锡的测定方法进行研究分析,结果表明:该方法检出限为0.002ng/mL,相对标准偏差(RSD)为3.05%~5.95%,加标回收率为94.0%~104.7%。使用原子荧光光谱法测定清洁水中微量锡具有灵敏度高、准确度、精密度好,仪器操作方便等优点,适宜水中微量锡的测定。 |
关键词: 氢化物原子荧光光谱法 水 微量锡 |
DOI: |
分类号: |
基金项目:浊度分光光发和目视比浊法标准修订项目(980) |
|
Determination of Micro-tin Content in Water by Atomic Fluorescence Spectrometry |
|
Abstract: |
In order to study detection of micro-tin content in water, AFS-820 Atomic Fluorescence Spectrometry was used. The results indicated that the detection limit was 0.002ng/mL. The relative standard deviation ranged from 1.49% to 5.95%, and the recovery ranged from 94.0% to 104.7%. The method has advantages of high sensitivity, accuracy, precision, easy equipment operating and etc., It is a good method to measure micro-tin content in water. |
Key words: atomic fluorescence spectrometry water micro-tin |